H2S gas monitor is equipped with 4-20mA and RS485 signal output, with a variety of relay alarm mode selection. Provides four selectable ranges for detecting hydrogen sulfide gas 0-20ppm, 0-50ppm, 0-100ppm, 0-200ppm, with a resolution of up to 0.01ppm or 0.1ppm. Combined with a high-brightness nixie tube display and 24-hour uninterrupted working capability, it is equipped with 4-20mA and RS485 signal outputs to provide all-round safety.
Specification
| Model |
GD300-H2S |
| Detection Principle |
Electrochemistry |
| Gas |
Hydrogen Sulfide (H2S) |
| Range (Optional) |
0 to 5ppm, 0 to 20ppm,, 0 to 50ppm, 0 to 100ppm, 0 to 200ppm |
| Resolution |
0.001ppm (0-5ppm), 0.01ppm (0-20/50ppm) or 0.1ppm (0-100/200ppm) |
| Calibration Kit |
Gas pressure regulator (calibration gas cylinder NOT included) |
| Accuracy |
≤±3% F.S. |
| Responce Time |
≤10s |
| Repeatability |
≤±1% |
| Recovery Time |
≤10s |
| Linearity Error |
≤±1% |
| Zero Shift |
≤±1% (F.S./year) |
| Signal Output (Optional) Note |
4-20mA and RS485 Relay output: one group of relay output (1NO+1NC, 5A/30V) for one alarm (default), or two groups of relay output (2NO+2NC, 5A/30V) for high/low alarm (optional). |
| Alarms |
Sound, light |
| Display |
High light nixie tube display |
| Installation |
Wall-mounted, piping type, flow-type (according to the detected environment) |
| Connection |
Power connection type: G1/2 male thread explosion-proof flexible pipe Pipe-mounted type: M45*P2.0 |
| Operating Time |
24h |
| Explosion-proof Grade |
II 2G Ex db IIC T4 Gb |
| Protection Grade |
IP65 |
| Certification |
CE, ATEX (Standard: EN IEC 60079-0:2018, EN 60079-1:2014) |
| Working Voltage |
12-36V DC |
| Case Material |
Die-casting aluminum housing |
| Operating Temperature |
-20℃~50℃ |
| Operating Humidity |
0~95%RH (non-condensing) |
| Dimension |
206*179*90mm |
| Weight |
≤1.6kg |
| Power Adapter |
Input 100-240VAC 50/60Hz, Output 24VDC 1A (Optional USA plug, Europe plug, AU plug or UK plug) |
Fixed Gas Monitor User Manual
Gas Monitor Functions
- This hydrogen sulfide gas detector adopts three-wire 4-20mA analog signal output, relay output, RS485 output (optional).
- The original imported sensors, with a unique infrared remote control function, non-contact operation and maintenance of hydrogen gas tester.
- Compatible with a variety of control alarms, PLC, DCS and other control systems, you can realize remote monitoring, gas monitoring is more accurate and reliable.
- Explosion-proof design, fast, credible and more stable.
- Full software auto-calibration function with temperature compensation can perfectly realize the compensation of gas concentration under different temperature environment.
Fixed H2S Gas Monitor Details

Fixed Hydrogen Sulfide Gas Monitor with Alarm Dimension (Unit: mm)

Tips: What is a fixed hydrogen sulphide (H2S) gas monitor?
A fixed hydrogen sulphide gas detector is a device used to monitor the concentration of hydrogen sulphide gas in the environment in real time. It is usually installed in a fixed location and is suitable for sites where continuous monitoring of gas leakage or concentration changes is required, such as industrial plants, mines and chemical laboratories. The instrument utilizes advanced sensing technology to provide accurate gas concentration data and is equipped with an alarm function to provide timely indication that the gas concentration is exceeding the limit to ensure personnel safety.